OPUS NITRATE SENSOR
OPUS NITRATE SENSOR

OPUS

OPUS is the new generation of spectral sensors for online measurement of nitrogen and carbon compounds. Through the analysis of a full spectrum, OPUS is able to deliver reliable readings for NO3-N, NO2-N, organic ingredients (CODeq, BODeq, DOCeq, TOCeq), and a…  Read more

Measurement technology

light source:

  • Xenon flash lamp

detector:

  • High-end miniature spectrometer
  • 256 Channels
  • 200 to 360 nm
  • 0.8 nm/pixel
Measurement principle Attenuation, spectral analysis
Optical path 0.3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 50 mm
Parameter See parameter list p. 3
Measuring range See parameter list p. 3
Measurement accuracy See parameter list p. 3
Turbidity compensation Yes
Data logger ~ 2 GB
T100 response time 2 min
Measurement interval ≥ 1 min
Housing material Stainless steel (1.4571/1.4404) or titanium (3.7035)
Dimensions (L x Ø)

470 mm x 48 mm (with 10 mm path)

~ 18.5“ x 1.9“ (with 10 mm path)
Weight

stainless steel: ~ 3 kg (with 10 mm path)

titanium: ~ 2 kg (with 10 mm path)

~ 6.6 lbs (with 10 mm path)

~ 4.4 lbs (with 10 mm path)

Interface

digital:

  • Ethernet (TCP/IP)
  • RS-232 or RS-485 (Modbus RTU)
Power consumption ≤ 8 W
Power supply 12...24 VDC (± 10 %)
Maintenance effort ≤ 0.5 h/month (typical)
Calibration/maintenance interval 24 months
System compatibility Modbus RTU
Warranty

1 year (EU: 2 years)

US: 2 years
Installation
Max. pressure

with SubConn: 30 bar

~ 435 psig
with fixed cable: 3 bar ~ 43.5 psig
in FlowCell: 1 bar, 2...4 L/min ~ 14.5 psig at 0.5 to 1 gpm
Protection type
  • IP68
  • NEMA 6P

Sample temperature

  • +2...+40 °C
  • ~ +36 °F to +104 °F
Ambient temperature
  • +2...+40 °C
  • ~ +36 °F to +104 °F
Storage temperature
  • -20...+80 °C
  • ~ -4 °F to +176 °F
Inflow velocity
  • 0.1...10 m/s
  • ~ 0.33 fps to 33 fps

  • Alt

OPUS is the new generation of spectral sensors for online measurement of nitrogen and carbon compounds. Through the analysis of a full spectrum, OPUS is able to deliver reliable readings for NO3-N, NO2-N, organic ingredients (CODeq, BODeq, DOCeq, TOCeq), and a number of other parameters. OPUS features the new TriOS G2 interface, allowing fast and easy configuration of sensors by using a web browser. Integration into existing process control systems and external
data loggers has never been easier. WiFi connectivity allows laptops, tablets or smartphones to be easily used for control without any special application software or app installation.

  • Without sampling and preparation of test samples
  • Real-time sensor
  • Without reagents
  • Optical window with nano coating
  • Pre-installed application calibration
  • Sewage treatment plants
  • Environmental monitoring
  • Drinking water monitoring
  • Industrial applications